MBW473_MANUAL_E_V2.0 5
2.4 Dew Point Measurement
When you switch on the 473, all readings on the display may initially be blank. If your instrument is
equipped with a pressure sensor, the pressure reading will be displayed. After connecting the measur-
ing head or temperature probe you will get an external temperature reading. To measure humidity
(dew point, frost point, %RH, etc.), the Dew/Frost Control mode must be enabled and gas must be
flowing across the mirror. If %RH readings are required, an external temperature probe must be con-
nected. To connect the external temperature probe, see section External Temperature Probe on
page 11. Alternatively, a fixed external temperature may be entered via the touch screen. Please fol-
low the instructions in section 5.2 External Temperature.
Fan
The SH2 measuring head is equipped with a fan to pull a gas sample across the
measuring head. If you are using the SH2 measuring head, press the Fan key on
the touch screen to turn the fan on/off. For further information on the fan refer to
section 3.3 Measuring Heads.